DI-718Bx "Save Default Setup" and "Open Reference File" Calibration Settings Error
On this Page
Symptoms
Cause
Resolution
Delete TPDATAQ.ini and re-install WinDaq software
Applies To
Symptoms
After selecting Save Default Setup from the File pull-down menu and restarting WinDaq, calibration settings are not retained.
Selecting Open Reference File from the File pull-down menu and choosing a previously recorded data file does not transfer calibration settings into WinDaq.
Cause
Software Bug
Resolution
To resolve these issues follow the steps below.
Delete TPDATAQ.ini and re-install WinDaq software
- Locate TPDATAQ.INI in the WINDOWS directory, and delete it.
- Download the latest installation from: www.dataq.com/support/downloads/new710.exe.
- Run the new installation software that you just downloaded, installing the application over top of the existing installation.
The software should now operate properly.
Applies To
All DI-718Bx Products.