Model EL-IOT-SP-TH
Wireless Data Logger Measures Temperatures from -30 to 80°C and Relative Humidity 0 to 100% RH
Stock
- Includes an EL-IOT-SP-TH WiFi data logger, a wall mounting bracket, 4 × 1.5V AA batteries, EL-SP-TH temperature and humidity probe and downloadable software (CALIBRATION CERTIFICATE sold separately under ACCESSORIES tab and must be ordered at time of purchase)
- Measures ambient temperature between -30 to +80 °C (-22 to +176 °F) and relative humidity 0 to 100% RH
- Records and uploads temperature data to the EasyLog Cloud
- Access live temperature data via any Internet Browser or the EasyLog Cloud App
- Set up alarm options for temperature zone breaches – email alerts, beeper alarms and/or LED flash alerts
- Typical battery life 6 months
- Quantity Discount
Description
The EL-IOT-SP-TH kit has been designed to monitor air temperature and relative humidity in a large range of applications. Its external temperature and humidity probe make it an ideal choice for applications where it may be problematic to locate the whole device such as museum display cases and HVAC systems.
Products in the EL-IOT range are simple to set-up and easy to use. Download the EasyLog Cloud App and set-up a device in minutes over your WiFi network. After set-up, the device can be placed anywhere within range of the WiFi network, continually monitoring and recording your data to the EasyLog Cloud. Access your data on any Internet Browser or the EasyLog Cloud App, enabling you to monitor the ambient temperature of your chosen location from anywhere at any time.
EL-IOT devices can be programmed with high and low temperature alarm zones. If a temperature zone is breached, an alarm will be activated. Alarm options include: email alerts (which can be sent to one or multiple email addresses), beeper sound alarms, and LED flash alerts.
EL-IOT devices can be powered using the supplied batteries or a mains AC power adapter (sold separately), and are provided with a wall mountable bracket for installation.
Accessories

Description
AC Power Adapter for EL- Data Loggers. Includes 4 interchangeable plug styles (UK, US, EU and AU/NZ). 1 meter cable length.

Description
2 meter Temperature and Humidity probe for EL-MOTE data loggers. Temperature Accuracy: ±0.3 (+5 to +60ºC) ±0.6 (+41 to +140ºF); Temperature Resolution: 0.1°C (0.1°F); Humidity Accuracy: ±2%RH (20 to 80%RH).
CAL EL-IOT-TEMPHUM-1
Calibration Certificate (1st Cal Value)

Description
NIST traceable calibration certificate (first point) in compliance with ISO 17025. Purchase this first point option first for each device, and then additional points per device as required. Standard points are $90. Custom points may be specified and entered into the special instructions section of our online store and incur an additional charge.
Please Note: Calibration certificates cannot be returned for a refund. These calibration certificates have a typical lead time of 3-4 weeks.
CAL EL-IOT-TEMPHUM-2
Calibration Certificate (2st Cal Value)

Description
NIST traceable calibration certificate (first point) in compliance with ISO 17025. Purchase this first point option first for each device, and then additional points per device as required. Standard points are $25. Custom points may be specified and entered into the special instructions section of our online store and incur an additional charge.
Please Note: Calibration certificates cannot be returned for a refund. These calibration certificates have a typical lead time of 3-4 weeks.
Software
EasyLog Cloud
Configuration Software
Cloud-based device management software and iOS and Android App for Lascar EL-Mote, EL-Wifi, El-IoT, and EL-WEM Data Loggers.

Specifications
Logger | |
Number of Channels: | One |
Battery Life: | 6 months (typical) Battery life is dependent on: transmission period, WiFi encryption method, WiFi encryption key rotation frequency (determined by the router/access point), signal strength between router/access point and WiFi device, presence, volume and type of WiFi traffic from other devices, sample rate and operating temperature. Logging period and transmission period can be configured in Settings via the EasyLog Cloud App |
Battery Type: | 4 × 1.5V AA cells |
Operating Temperature (Mains Adapter Powered): | -20 to +60°C (-4 to +140°F) |
Operating Temperature (Powered by Batteries Supplied): | -18 to +55°C (0 to +131°F) |
Logging Rate: | 1 sample every 10 seconds 1 sample every minute 1 sample every 5 minutes 1 sample every 30 minutes 1 sample every hour 1 sample every 6 hours 1 sample every 12 hours |
Transmission Period (User Configurable): | 1 minute 2 minutes 5 minutes 10 minutes 15 minutes 30 minutes 1 hour 1.5 hours 2 hours 3 hours 4 hours 5 hours 6 hours 12 hours 24 hours |
Dimensions: | 94 × 83 × 31 mm (3.7 × 3.3 × 1.2 inches) |
Environmental Rating: | IP54 (Splash proof) |
Probe(s) | |
Temperature Resolution: | 0.1° |
Temperature Accuracy: | Typical: ±0.3°C (±0.6°F) (+5 to +60°C / +41 to +140°F) Maximum: ±1.2°C (±2.2°F) (-20 to +80°C / -4 to +176°F) |
Humidity Measurement Range: | 0 to 100% RH |
Humidity Resolution: | 1% RH |
Humidity Accuracy (@25˚C): | Typical: ±2% RH (20% to 80% RH) Maximum: ±5% RH (0% to 100% RH) |
Cable Length: | 2 m (79 inches) |